Switching the centrifuge on and off.

Switching the centrifuge on
AB4701en
NOTICE
Danger due to condensation in the event of temperature
differences
Moisture may damage electrical components.
Ensure that all surfaces are dry prior to commissioning
or maintenance.
If the temperature changes, wait until the device or com-
ponent has become acclimatised.
Prevent moisture from penetrating into sensitive compo-
nents.
If moisture forms, switch off the device immediately and
allow it to dry properly.
Personnel:
Trained user
1.
A type B residual current circuit breaker must be used if the device
is additionally protected with a residual current circuit breaker in the
building installation.
When using a different type, the residual current circuit breaker may
either not switch off the unit if there is a fault on the unit, or it may
switch off the unit even though there is no fault on the unit.
2.
Centrifuge with serial interface:
Connect the centrifuge's RS232 interface to the PC using an RS232
connection cable.
3.
Centrifuge with PE connector:
If necessary, connect the PE connector on the back of the device to
an additional medical potential equalisation system.
4.
For the centrifuge, order no. 4706-20: connect the nitrogen supply
according to the enclosed instruction sheet AH4706-20XX.
5.
Check whether the mains voltage and mains frequency match the
specification on the rating plate.
6.
Connect the device to a standard mains socket using the mains
cable.
Personnel:
Trained user
Rev.: 19 / 02.2025
Commissioning
25 / 64