Operation

Immediate display of centrifu-
gation data after switching on
Switching off the centrifuge
6
Operation
6.1

Opening and closing the lid

Opening the lid
Closing the lid
26 / 64
Set the mains switch to
�� The buttons flash, depending on the centrifuge type.
The following indicators appear one after the other, depending on
the centrifuge type:
the centrifuge model
the program version and mains voltage
the rotor code (Rotor), the maximum rotor speed (Nmax) and
the centrifuging radius (R) of the last rotor recognised by the
rotor detection.
The centrifuging radius displayed is a default value that must be
adjusted depending on the accessories used.
When the lid is closed:
When the lid is open: The centrifugation data of the last pro-
gram used or program 1.
1.
Set the mains switch to
2.
Press and hold any button at the first visual change in the display
(inverse display).
�� Centrifugation data is displayed.
The rotor is stationary.
Set the mains switch to
Personnel:
Trained user
The centrifuge is switched on
The rotor is stationary.
Press the
[STOP/OPEN] button.
�� The lid unlocks by means of a motor.
The light on the left side of the
CAUTION
Crushing hazard when closing the lid.
Danger of fingers getting crushed when the closing motor
pulls the lid against the seal.
No parts of the operator's body should be in the hazard
zone of the lid when closing the lid.
To close the lid, press on the lid from above.
NOTICE
Damage to the device caused by the lid slamming.
Close the lid slowly.
Do not slam the lid.
Rev.: 19 / 02.2025