Instructions For Cleaning And Disinfection - Hettich Rotina 420 R Mode D'emploi

Masquer les pouces Voir aussi pour ROTINA 420 R:
Table des Matières
Cleaning and care
Chap.
Task to execute
8.3
Cleaning accessories
8.4
Disinfection
8.4
Disinfecting the device
8.4
Disinfecting the accessories
8.5
Maintenance
8.5
Greasing the rubber seal of the centrifuging
chamber
8.5
Greasing the rubber seal of the biosafety system
8.5
Trunnion greasing
8.5
Checking the accessories
8.5
Checking the biosafety system
8.5
Centrifuging chamber damage inspection
8.5
Greasing the motor shaft
8.5
Accessories with a limited service life
8.5
Replacing centrifuge tubes
8.2

Instructions for cleaning and disinfection

54 / 64
DANGER
Risk of contamination
Inadequate cleaning or failure to observe the cleaning
instructions can lead to contamination risks.
Observe national and local regulations on safety and
accident prevention.
Observe cleaning instructions.
Wear personal protective equipment when cleaning the
device.
Comply with laboratory regulations (e.g. TRBAs, the
German Protection against Infection Act, hygiene plan)
for handling biological agents.
The device and its accessories must not be cleaned in dishwashers.
Only perform hand cleaning and liquid disinfection.
The water temperature must not exceed 25 °C.
To prevent any corrosion due to use of detergents or disinfectants, it is
essential to follow the special application instructions provided by the
manufacturers of the detergent or disinfectant.
Disinfectant:
Rev.: 19 / 02.2025
 
 
X
 
 
 
X
 
 
X
 
 
 
 
 
 
 
X
 
 
X
 
 
X
 
 
X
 
 
X
 
 
 
 
 
 
X
 
 
X
 
 
 
55
 
55
 
56
 
56
 
56
 
56
 
56
 
56
 
57
 
57
X
57
X
57
 
57
 
57
AB4701en