Switching the centrifuge on and off.

Switching the centrifuge on
AB5650en
1
Nitrogen outlet
2
Nitrogen inlet, throttle valve
2.
The nitrogen inlet (
connected via a pneumatic quick-release fitting with a 6 mm pneu-
matic hose.
3.
The nitrogen outlet (
mm hose connection. Attach the hose onto the screw-in fitting.
4.
Check whether the mains voltage matches the specification on the
rating plate.
5.
Connect the device to a standard mains socket using the mains
cable.
Monitoring the nitrogen flow
The operator is responsible for monitoring the nitrogen flow.
Technical conditions
Nitrogen inlet: pressure upstream of the throttle valve, 0.4 bar.
Personnel:
Trained user
Set the mains switch to
�� The buttons flash, depending on the centrifuge type.
The following indicators appear one after the other, depending on
the centrifuge type:
the centrifuge model
the program version and mains voltage
the rotor code (Rotor), the maximum rotor speed (Nmax) and
the centrifuging radius (R) of the last rotor recognised by the
rotor detection.
The centrifuging radius displayed is a default value that must be
adjusted depending on the accessories used.
When the lid is closed:
When the lid is open: The centrifugation data of the last pro-
gram used or program 1.
Rev.: 24 / 01.2024
2 ) is located at the back of the machine and is
1 ) is located on the lid and is connected via a 12
[I] .
'OPEN OEFFNEN' indicator
Commissioning
1
2
31 / 73