Enabling or disabling integral
RCF
7.1.5
Temperature (for centrifuges with cooling)
7.1.6
Relative centrifugal force, RCF
AB5650en
[RCF] button repeatedly until the integral RCF is displayed.
1.
Press the
[RCF] button.
2.
Press the
The centrifugation data is displayed.
3.
If necessary, press the
The RPM is displayed.
1.
Press and hold the
'***Machine Menu***' is displayed after 8 seconds.
[PROG] button repeatedly until '-> Settings' is displayed.
2.
Press the
[START] button.
3.
Press the
'SOUND / BELL = on' or 'SOUND / BELL = off' is displayed.
[PROG] button repeatedly until 'RCF Integral = on' or 'RCF
4.
Press the
Integral = off' is displayed.
[Rotary knob] to set 'off' or 'on' .
5.
Use
off = integral RCF disabled
on = integral RCF enabled.
[START] button.
6.
Press the
The setting is stored.
'Store Settings ...' is displayed briefly
'-> Settings' is then displayed.
[OPEN/STOP] button once to exit the 'Settings menu'
7.
Press the
or
[OPEN/STOP] button twice to exit the 'Machine Menu' .
Press the
[T/°C] button.
1.
Press the
The parameter T/°C or T/°F is displayed.
2.
Use the rotary knob to set the desired value.
[T/°C] button or the [START] button.
3.
Press the
The setting is transferred to the indicator.
The relative centrifugal force RCF is dependent on the speed and the centri-
fuging radius.
The relative centrifugal force RCF is stated as a multiple of the acceleration
due to gravity (g).
The relative centrifugal force RCF is a dimensionless numerical value and is
used to compare the separation and sedimentation performance.
RCF = Relative Centrifugal Force
RPM = speed
Rev.: 24 / 01.2024
[RPM] button.
[PROG] button.
Software operation