Operation
Removing the bucket
6.4

Inserting and removing adapters

Inserting
the adapter
removing
6.5

Loading

Filling centrifuge tubes
34 / 73
1
2
3
1.
Check that the rotor is firmly seated.
2.
Grease the trunnions (3).
3.
Insert the bucket (1) into the rotor from above. The trunnions (3) must
be in the grooves (2).
4.
Push the bucket (1) down as far as it will go.
Pull the bucket (1) vertically upwards out of the rotor.
1
2
Insert the adapter (1) vertically into the bucket (2) from above.
Remove the adapter (1) vertically upwards out of the bucket (2).
WARNING
Risk of injury from contaminated sample material.
Contaminated sample material escapes from the sample
tube during centrifugation.
Use centrifuge tubes with special screw caps for haz-
ardous substances.
For risk group 3 and 4 materials, use a biosafety system
in addition to the sealable centrifuge tubes (see WHO's
'Laboratory Biosafety Manual').
Rev.: 24 / 01.2024
1
2
3
1
2
AB5650en