Software operation
7.4.6

Temperature monitoring

7.5

Heating (for centrifuges with heating)

50 / 73
[PROG] button repeatedly until 'Cool dec speed = ... rpm' is
4.
Press the
displayed.
[Rotary knob] to set the desired value.
5.
Use the
[START] button.
6.
Press the
�� The setting is stored.
'Store Settings...' is displayed briefly.
'-> Settings' is then displayed.
[STOP/OPEN] button once to exit the 'Settings menu'
7.
Press the
or
[STOP/OPEN] button twice to exit the '***Machine Menu***' .
Press the
Temperature monitoring is used to protect temperature-sensitive samples.
The temperature is monitored after reaching the set temperature range. The
target temperature range is set to target temperature ±3°C.
The centrifugation run is aborted and the error message
58.6' is displayed if the temperature in the centrifuging chamber exceeds
the set temperature by the value
The centrifugation run is aborted and the error message
58.7' is displayed if the temperature in the centrifuging chamber falls below
the set temperature by the value
1.
Press and hold the
'***Machine Menu***' is displayed after 8 seconds.
��
[PROG] button repeatedly until '-> Settings' is displayed.
2.
Press the
3.
Press the
[START] button.
'SOUND / BELL = on' or 'SOUND / BELL = off' is displayed.
��
[PROG] button repeatedly until 'Error 58 Temp 15 °C' is
4.
Press the
displayed.
[Rotary knob] to set the desired value.
5.
Use the
Adjustable from 4 °C to 25 °C, in 1 °C increments, as well as the "dis-
abled" setting. Temperature monitoring is deactivated when "disabled"
is set.
[START] button.
6.
Press the
�� The setting is stored.
'Store Settings...' is displayed briefly.
'-> Settings' is then displayed.
[STOP/OPEN] button once to exit the 'Settings menu'
7.
Press the
or
[STOP/OPEN] button twice to exit the '***Machine Menu***' .
Press the
During the centrifugation run, the centrifuging chamber is heated to the
preselected temperature if required. The heating is switched off when the
rotor is at a standstill.
Swing-out rotors and angle rotors must run at maximum speed.
Rev.: 24 / 01.2024
'Error 58 Temp' for more than 2 minutes.
'Error 58 Temp' for more than 2 minutes.
[PROG] button.
'°C/ * -ERROR
'°C/ * -ERROR
AB5650en