Software operation
7.6.1.1
Centrifuge address
7.6.2
Cycle counter
Entering the maximum per-
missible number of run cycles
52 / 73
[PROG] button.
5.
Press the
The rotor code (Rotor), the maximum rotor speed (Nmax) and the
centrifuging radius (R) of the last rotor recognised by the rotor
detection are displayed.
The last rotor recognised is marked with an asterisk (*).
[Rotary knob] can be used to display the information of the
The
rotors approved for the centrifuge.
[PROG] button.
6.
Press the
The program version of the centrifuge is displayed.
[PROG] button.
7.
Press the
The program version of the frequency converter is displayed.
[STOP/OPEN] button twice to exit the '-> Info' menu
8.
Press the
or
[STOP/OPEN] button three times to exit the '***Machine
Press the
Menu***' .
The address of the centrifuge is set to ]=29 ex works. Address set.
The centrifuge is equipped with a cycle counter. The cycle counter counts
the run cycles (centrifugation cycles) of the various rotor codes.
For swing-out rotors, the cycle counter is used to record the run cycles
(centrifugation runs) of the buckets.
The centrifugation run is cancelled when the rotor is recognised by the
rotor detection for the first time.
after pressing any button. The maximum permissible number of run cycles
indicated on the bucket must be entered before the centrifugation run can
be restarted.
The cycle counter can be disabled for rotors and buckets that are not
marked with the maximum permissible number of run cycles. The number
of run cycles (centrifugation runs) of the rotor code used is displayed briefly
every time the lid is opened.
If the maximum permissible number of bucket run cycles entered is
'*MAX CYCLES PASSED*' is displayed after each start of a
exceeded,
centrifugation run.
The centrifugation run must be restarted. The buckets must be replaced
with new ones.
If the buckets have been replaced, the cycle counter must be reset to
The maximum permissible number of run cycles must be entered after
starting the first centrifugation run.
'Enter max cycles = á30000 ñ' is displayed.
[Rotary knob] to set the maximum permissible number of run
1.
Use the
cycles indicated on the bucket.
2.
Press the
[START] button.
The setting is stored.
'Store max cycles ...' is displayed briefly.
Rev.: 24 / 01.2024
'Enter max cycles = á30000 ñ' is displayed
'0' .