Machine Menu; Querying System Information - Hettich Rotanta 460 R Mode D'emploi

Masquer les pouces Voir aussi pour ROTANTA 460 R:
Table des Matières
Les langues disponibles

Les langues disponibles

enable / disable
7.6

Machine Menu

7.6.1

Querying system information

AB5650en
CAUTION
Danger of burns from hot surfaces.
The surface temperature of the heating element in the centri-
fuging chamber can be up to 500 °C or 932 °F.
Do not touch the heating element.
NOTICE
Damage to plastic buckets due to excessive temperature
Plastic buckets may only be used at temperatures up to
a maximum of 40 °C or 104 °F.
To achieve an exact temperature, a pre-tempering run of up
to 60 minutes must be carried out before each centrifuga-
tion run.
The rotor is stationary.
[T/°C] button repeatedly until 'Heater = off' or 'Heater = on'
1.
Press the
is displayed.
[Rotary knob] to set 'off' or 'on' .
2.
Use
off = heating disabled
on = heating enabled
[T/°C] button or the [START] button.
3.
Press the
�� The settings are stored.
Centrifugation data is displayed.
The following system information can be queried:
Centrifuge model
Mains voltage
Rotor information
Centrifuge program version
Program version for the frequency inverter
The rotor is stationary.
1.
Press and hold the
'***Machine Menu***' is displayed after 8 seconds.
��
[PROG] button repeatedly until '-> Info' is displayed.
2.
Press the
[START] button.
3.
Press the
�� The centrifuge model is displayed.
[PROG] button.
4.
Press the
�� The mains voltage is displayed
Rev.: 24 / 01.2024
[PROG] button.
Software operation
51 / 73
Table des Matières
loading

Ce manuel est également adapté pour:

Rotanta 460Rotanta 460 rcRotanta 460 rf