Brake level and ramp-down
time
Brake cut-off speed
7.1.2
Runtime TIME
Changing the runtime
AB5650en
4.
If required: Press the
to set the next parameter.
[START] button.
5.
Press the
or
[Ramp-up and ramp-down parameters] button repeatedly
Press the
until the centrifugation data is displayed.
B-brake levels can only be set for rotors that are suitable for
use with blood bags.
−
Setting the B-brake levels is only possible if they are
enabled.
−
Setting ramp-down times is only possible if they are
enabled.
'Ramp-down time' function is enabled.
The
[Ramp-up and ramp-down parameters] button repeatedly
1.
Press the
'Brake level' parameter, 'B-brake level' or the 'Ramp-down
until the
time' parameter is displayed.
2.
Press the
[TIME] button to toggle between the brake level and the
ramp-down time.
[Rotary knob] to set the desired level or time.
3.
Use the
4.
If required: Press the
to set the next parameter.
[START] button.
5.
Press the
or
[Ramp-up and ramp-down parameters] button repeatedly
Press the
until the centrifugation data is displayed.
[Ramp-up and ramp-down parameters] button repeatedly
1.
Press the
'N Brake' parameter is displayed.
until the
[Rotary knob] to set the desired value.
2.
Use the
[Ramp-up and ramp-down parameters] button
3.
The
or
[START] button.
Press the
Settings are shown in the indicator.
For continuous operation, the minutes, seconds and hours
must be set to zero.
Continuous operation is indicated in the indicator by means
of the '¥' symbol.
[TIME] button.
1.
Press the
't/hms' is displayed.
The minutes are displayed in brackets á ñ.
2.
Use the
[Rotary knob] to set the desired value.
Rev.: 24 / 01.2024
[Ramp-up and ramp-down parameters] button
[Ramp-up and ramp-down parameters] button
Software operation
43 / 73