Insert the blood bag (1) into the insert (2).
1
2
2.
Hold the blood bag by the connectors (1) and push the support plate
(2) on the outer side of the blood bag into the insert from top to
bottom.
Ensure that the lower edge of the support plate stands on the floor as
completely as possible.
1
3.
Fold the support plate outwards and press it down until the folded
edge of the support plate is level with the liquid level of the blood bag.
The upper edge of the support plate must not project too far out of
the insert during centrifugation due to the risk of its becoming caught
in the rotor arms.
Note the position of the loop (1) so that it can be reached after
centrifugation.