5.3

Setting up and connecting the centrifuge

Setting up the centrifuge
AB2200en
3
2
3
2
1
2 3
Fig. 14: Transport lock
1
Transport lock
2
Screws
3
Spacer sleeves
4
Right side of device
5
Bellows
1.
Lay the device down onto its right-hand side (
2.
Remove 3 screws (
3.
Remove 3 transport locks (
4.
Keep the screws, spacer sleeves and transport locks in a safe place.
5.
For MIKRO 220 R:
Open the lid.
Check the bellows (
seating.
WARNING
Risk of injury due to failing to maintain a sufficient distance to
the centrifuge.
As per EN / IEC 61010-2-020, no persons, hazardous
materials or objects may be present within a safety zone
of 300 mm around the centrifuge during a centrifugation
run.
A distance of 300 mm from the ventilation slots and ven-
tilation openings of the centrifuge must be maintained.
CAUTION
Risk of crushing and damage to the device due to it falling
down because of vibration-induced position alterations.
Place the device on a stable and level surface.
Select the installation surface dependent on the weight
of the device.
Rev.: 15 / 01.2024
5
4
2  ) and 3 spacer sleeves ( 3  ).
1  ).
5  ) underneath the motor cover for correct
Commissioning
1
3
2
4  ).