Software operation
7.2
Programming
7.2.1
Write protection for programs
7.2.2
Opening or loading programs
7.2.3
Entering or changing programs
28 / 39
If, in exceptional cases, the maximum load indicated on the bucket is
exceeded, the speed must also be reduced. The permissible speed can
be calculated using the following formula:
For example: Maximum speed 4000 RPM, maximum load 300 g, actual
load 350 g
Please contact the manufacturer if you are not sure.
The programs can be protected to prevent unintentional changes.
When the rotor is at a standstill, write protection can be enabled or disabled
as follows:
1.
Press and hold the
'SOUND/BELL' is displayed after 8 seconds.
[SELECT] button.
2.
Press the
'LOCK' is displayed.
[Rotary knob] to set 'OFF' or 'ON' .
3.
Use
OFF = The programs are not write-protected
ON = The programs are write-protected
[START/IMPULS] button.
4.
Press the
The setting is stored.
'*** lock ***' is displayed briefly.
If ON is set:
'*** ok ***' is displayed briefly.
If OFF is set:
[SELECT] button to select the 'PROG RCL' parameter.
1.
Use the
[Rotary knob] to set the desired program location.
2.
Use the
[START/IMPULS] button.
3.
Press the
'*** ok ***' is displayed briefly.
The centrifugation data of the desired program location is dis-
played
4.
To check the parameters: Press the
5.
To exit the parameter indicator: Press the
not press any button for 8 seconds.
1.
Retrieve program.
2.
If required: Press the
'> <' ).
indicator (
Rev.: 15 / 01.2024
[SELECT] button.
[SELECT] button several times.
[OPEN/STOP] button or do
[RCF] button to toggle between RPM and RCF
AB2200en