Télécharger Imprimer la page

avanos MIC Mode D'emploi page 3

Masquer les pouces Voir aussi pour MIC:

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 5
Dilation
1. Use a #11 scalpel blade to create a skin incision that extends alongside the guidewire, downward through the
subcutaneous tissue and fascia of the abdominal musculature. After the incision is made, dispose of the scalpel according
to facility protocol.
2. Advance a dilator over the guidewire and dilate the stoma tract at least four French sizes larger than the enteral feeding
tube being placed.
3. Remove the dilator over the guidewire, leaving the guidewire in place.
Note. After dilation, a peel-away sheath may be used to facilitate advancement of the tube through the stoma tract.
Tube Placement under Fluoroscopic (Radiologic) Visualization
1. Advance the distal end of the tube over the guidewire, through the stoma tract and into the stomach.
2. Rotate the MIC* GJ Feeding Tube while advancing to facilitate passage of the tube through the pylorus and in the
jejunum.
3. Advance the tube until the tip of the tube is 10–15 cm beyond the Ligament of Treitz and the balloon is in the stomach.
4. Using the male Luer syringe, inflate the balloon with 7–10 ml of sterile or distilled water.
Caution. Do not exceed 20 ml total balloon. Do not use air. Do not inject contrast into the balloon.
5. Gently pull the tube up and away from the abdomen until the balloon contacts the inner stomach wall.
6. Clean the residual fluid or lubricant from the tube and stoma.
7. Gently slide the external retention bolster to approximately 2–3 mm above the skin. Do not suture the bolster to the skin.
8. Remove the guidewire.
Tube Placement under Endoscopic Visualization
1. Advance the distal end of the tube over the guidewire, through the stoma tract and into the stomach.
2. Grasp the suture loop or the tip of the tube with atraumatic forceps.
3. Advance the MIC* GJ Feeding Tube through the pylorus and upper duodenum. Continue to advance the tube using the
forceps until the tip is positioned 10-15 cm beyond the Ligament of Treitz and the balloon is in the stomach.
4. Release the tube and withdraw the endoscope and forceps in tandem, leaving the tube in place.
5. Using the male Luer syringe, inflate the balloon with 7–10 ml of sterile or distilled water.
Caution. Do not exceed 20 ml total balloon. Do not use air. Do not inject contrast into the balloon.
6. Gently pull the tube up and away from the abdomen until the balloon contacts the inner stomach wall.
7. Clean the residual fluid or lubricant from the tube and stoma.
8. Gently slide the external retention bolster to approximately 2–3 mm above the skin. Do not suture the bolster to the skin.
9. Remove the guidewire.
Verify Tube Position and Patency
1. Verify proper tube placement radiologically to avoid potential complication (e.g. bowel irritation or perforation) and
ensure the tube is not looped within the stomach or small bowel.
Caution: The jejunal portion of the tube contains tungsten, which is radiopaque and can be used to radiographically
confirm position. Do not inject contrast into the balloon.
2. Flush both the jejunal and gastric lumens with water to verify patency.
3. Check for moisture around the stoma. If there are signs of gastric leakage, check the tube position and placement of the
external retention bolster. Add sterile or distilled water as needed in 1–2 ml increments. Do not exceed balloon capacity
as indicated previously.
4. Check to ensure that the external retention bolster is not placed too tightly against the skin and rests 2–3 mm above the
abdomen for initial placement and 1–2 mm above the abdomen for a replacement tube.
5. Begin feeding only after confirmation of proper patency, placement and according to physician instructions.
Replacement Procedure through an Established Gastrostomy Tract
1. Cleanse the skin around the stoma site and allow the area to air dry.
2. Select the appropriate size MIC* GJ Feeding Tube and prepare according to the instructions in the Tube Preparation
section above.
3. If using endoscopic visualization perform routine esophagogastroduodenoscopy (EGD). Once the procedure is complete
and no abnormalities are identified that could pose a contraindication to placement of the tube, place the patient in the
supine position and insufflate the stomach with air. Manipulate the endoscope until the indwelling gastrostomy tube is
in the visual field.
4. Under either fluoroscopic guidance or endoscopic guidance, insert a floppy-tipped guidewire, up to .038", through the
indwelling gastrostomy tube.
5. Remove the gastrostomy tube over the guidewire.
6. Place the MIC* GJ Feeding Tube according to the Tube Placement section above.
Caution: For a replacement feeding tube, gently slide the external retention bolster to approximately 1–2 mm above
the skin. Do not suture the bolster to the skin.
7. Verify tube position and patency using the Verify Tube Position and Patency section above.
Tube Patency Guidelines
Proper tube flushing is the best way to avoid clogging and maintain tube patency. The following are guidelines to avoid
clogging and maintain tube patency.
Flush the feeding tube with water every 4–6 hours during continuous feeding, anytime the feeding is interrupted, or at
least every 8 hours if the tube is not being used.
Flush the feeding tube after checking gastric residuals.
Flush the feeding tube before and after medication administration and between medications. This will prevent the
medication from interacting with formula and potentially causing the tube to clog.
Flush the feeding tube after checking gastric residuals.
Use liquid medication when possible and consult the pharmacist to determine if it is safe to crush solid medication and
to mix with water. If safe, pulverize the solid medication into a fine powder form and dissolve the powder in warm water
before administering through the feeding tube. Never crush enteric-coated medication or mix medication with formula.
Avoid using acidic fluids such as cranberry juice and cola beverages to flush feeding tubes as the acidic quality when
combined with formula proteins may actually contribute to tube clogging.
General Flushing Guidelines
Flush the feeding tube with water using an ENFit® syringe every 4–6 hours during continuous feeding, anytime the feeding
is interrupted, at least every 8 hours if the tube is not being used, or per clinician's instructions. Flush the feeding tube after
checking gastric residuals. Flush the feeding tube before, after, and in between medication administration. Avoid using acidic
irrigants such as cranberry juice and cola beverages to flush feeding tubes.
Use a 30 to 60 ml ENFit® syringe. Do not use smaller size syringes as this can increase pressure on the tube and potentially
rupture smaller tubes.
Use room temperature water for tube flushing. Sterile water may be appropriate where the quality of municipal water
supplies is of concern. The amount of water will depend on the patient's needs, clinical condition, and type of tube,
but the average volume ranges from 10 to 50 ml for adults, and 3 to 10 ml for infants. Hydration status also influences
the volume used for flushing feeding tubes. In many cases, increasing the flushing volume can avoid the need for
supplemental intravenous fluid. However, individuals with renal failure and other fluid restrictions should receive the
minimum flushing volume necessary to maintain patency.
Do not use excessive force to flush the tube. Excessive force can perforate the tube and can cause injury to the
gastrointestinal tract.
Document the time and amount of water used in the patient's record. This will enable all caregivers to monitor the
patient's needs more accurately.
Nutrition Administration
1. Open the cap to the Jejunal access port of the MIC* GJ Feeding Tube.
2. Use an ENFit® syringe to flush the tube with the prescribed amount of water as described in the General Flushing
Guidelines.
3. Remove the flushing syringe from the Jejunal access port.
4. Securely connect an ENFit® feed set to the Jejunal access port.
Caution: Do not over-tighten the feed set connector or the syringe to the access port.
5. Complete feeding per the clinician's instructions.
Warning: If formula is present in the gastric drainage, stop feeding and notify the physician or health
care provider.
6. Remove the feed set or syringe from the Jejunal access port.
7. Use an ENFit® syringe to flush the tube with the prescribed amount of water as described in the General Flushing
Guidelines.
8. Remove the flushing syringe from the Jejunal access port.
9. Close the cap to the Jejunal access port.
Medication Administration
Use liquid medication when possible and consult the pharmacist to determine if it is safe to crush solid medication and mix
with water. If safe, pulverize the solid medication into a fine powder form and dissolve the powder in warm water before
administering through the feeding tube. Never crush enteric coated medication or mix medication with formula.
1. Open the cap to the prescribed access port of the MIC* GJ Tube.
2. Use an ENFit® syringe to flush the tube with the prescribed amount of water as described in the General Flushing
Guidelines.
3. Remove the flushing syringe from the access port.
4. Securely connect an ENFit® syringe containing the medication to the access port.
Caution: Do not over-tighten the syringe to the access port.
5. Deliver the medication by depressing the ENFit® syringe plunger.
6. Remove the syringe from the access port.
7. Use an ENFit® syringe to flush the tube with the prescribed amount of water as described in the General Flushing
Guidelines.
8. Remove the flushing syringe from the access port.
9. Close the cap to the access port.
Gastric Decompression
Gastric decompression may be performed via either gravity drainage or low intermittent suction.
Caution: Never connect the Jejunal access port to suction. Do not measure residuals from the Jejunal access port.
1. Open the cap to the Gastric access port of the MIC* GJ Feeding Tube.
2. For gravity drainage, place the opened Gastric access port of the MIC* GJ Feeding Tube directly over the opening of an
appropriate container.
Note: Ensure the open access port is positioned below the stoma.
3. For low intermittent suction, connect an ENFit® syringe to the Gastric access port.
4. Apply low intermittent suction by slowly retracting the plunger of the syringe in short intervals.
Warning: Do not use continuous or high pressure suction. High pressure could collapse the tube or injure
the stomach tissue and cause bleeding.
5. Disconnect the decompression syringe from the Gastric access port.
6. Use an ENFit® syringe to flush the tube with the prescribed amount of water as described In the General Flushing
Guidelines.
7. Remove the flushing syringe from the Gastric access port.
8. Close the cap to the Gastric access port.
Balloon Maintenance
Precise balloon life cannot be predicted. Silicone balloons generally last 1–8 months, but the life span of the balloon varies
according to several factors. These factors may include medications, volume of water used to inflate the balloon, gastric pH,
and tube care.
Check the water volume in the balloon once a week.
Insert a male Luer syringe into the Balloon Inflation Port (BAL.) and withdraw the fluid while holding the tube in place.
Compare the amount of water in the syringe to the amount recommended or the amount initially prescribed and
documented in the patient record. If the amount is less than recommended or prescribed, refill the balloon with the
water initially removed, then draw up and add the amount needed to bring the balloon volume up to the recommended
and prescribed amount of water. Be aware as you deflate the balloon there may be some gastric contents that can leak
from around the tube. Document the fluid volume, the amount of volume to be replaced (if any), the date and time.
Wait 10–20 minutes and repeat the procedure. The balloon is leaking if it has lost fluid, and the tube should be replaced.
A deflated or ruptured balloon could cause the tube to dislodge or be displaced. If the balloon is ruptured, it will need
to be replaced. Secure the tube into position using tape, then follow facility protocol and/or call the physician for
instructions.
Caution: Refill the balloon using sterile or distilled water, not air or saline. Saline can crystallize and clog the balloon
valve or lumen, and air may seep out and cause the balloon to collapse. Be sure to use the recommended amount of water
as over-inflation can obstruct the lumen or decrease balloon life and under-inflation will not secure the tube properly.
Daily Care & Maintenance Check List
Assess the patient for any signs of pain, pressure or discomfort.
3

Publicité

loading

Produits Connexes pour avanos MIC