18.1 Error messages

Error message Explication
ERR H
Hardware error
High zero point shift during
adjustment
ERR C
Objects present on weighing
plate.
Missing weighing plate
CAL D
Display instable
Error at numerical input (e. g.
ERR N
wrong password)
ERR W
Wrong application
COM ERR
Wrong remote control order.
- OL
Missing weighing plate
OL
Overload
ABORT
Process cancelled
WAIT
Wait for process
When adjustment was started
BUSY
objects were present on
weighing plate
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
96
Remedy
Switch balance off and on
again. If the error message
remains inform manufacturer.
Use ON/OFF to return into
weighing mode. Restart
adjustment process.
Check ambient conditions
(draught, vibrations etc.) Use
ON/OFF to return into weighing
mode. Restart adjustment
process.
Correct input
The balance changes to the
previous state. Correct
application.
The balance changes to the
previous state. Correct remote
control order.
Install weighing plate correctly
Reduce load
Remove object and continue
adjustment