Zero / Tare Menu (Zeroing And Taring Functions) - KERN and SOHN ABS-N Série Mode D'emploi

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

12 Zero / tare menu (zeroing and taring functions)

Selectable functions:
1. Zero tracking function
+
see chpt 12.1
2. Auto zero function
+
see chpt 12.2
3. Auto tare function
+
see chpt 12.3
4. Zero / tare timing
change function
+
see chpt 12.4
48
This function is used to correct automatically small
weight variations which appear directly after
switching-on.
In the event that small quantities are removed or added
to the material to be weighed, incorrect weighing results
can be displayed due to the "stability compensation" in
the balance. (e.g. slow flow of liquids from a container
placed on the balance, evaporating processes).
When apportioning involves small variations of weight, it
is advisable to switch off this function.
This function is used to correct automatically small
weight variations which appear after a measurement
(e.g. soiled weighing plate) after the stability display.
After data output an automatic taring is carried out
Selectable if the balance tares or gets set to zero
before or after indication of the stability mark.