12.2 Auto zero function
The auto zero function is not available when formula mode is enabled
(see chapter 14.3).
1. Call up menu
Press TARE for 3 sec. in weighing mode.
2. Select function
Press the navigation buttons ( ) repeatedly until
„A.ZERO" is displayed. The current setting is marked by
the stability display ( ).
Further steps depend on the desired setting:
Stability
Function
display ( )
yes
ON
no
off
3. Define zero range
Press TARE
Use the navigation keys ( ) to enter the zero range
see chpt. 3.1.1 „Numeric input".
4. Return to weighing mode
Press ON/OFF repeatedly or for 3 sec.
When entering the zero range observe the currently set weighing unit.
If the weighing unit is changed at a later moment, the zero range must be
adapted to the new weighing unit (step 3.).
Upper limit zero range: 99 d (in the displayed weighing unit)
Lower limit zero range: 1d (in the displayed weighing unit)
Example for a balance d = 0.0001 g
50
set/update
Press PRINT and
continue with step 3.
Press TARE and
continue with step 3.
Unit
Lower limit
g
0.0001 g
ct
0.001 ct
cancel
Press TARE and
continue with step 4.
continue with step 4.
Upper limit
0.0099 g
0.099 ct
ABS-N_ABJ-NM-BA-e-1614