Or count more pieces
or
The menu item to change the stored reference is called up, the
currently set reference quantity is displayed.
+
see „4. Change or add piece weight"
The stored piece weight is displayed in grams, marked by the
symbol
can be edited when a printer is connected.
e.g.: UW =
To go back to piece counting display press UNIT anew.
Switching over between piece counting and weighing mode
4. Change or add piece weight
In piece counting mode call-up the desired memory
location
Every time UNIT is pressed (approx. for 3 sec.), the next
memory location [
Change or add piece weight
Place an empty container on the balance and tare it by
pressing TARE.
Press MENU twice, the currently set reference quantity
is displayed.
Press the navigation buttons ( ) repeatedly until the
desired reference quantity is displayed (possible
reference quantities 5, 10, 20, 50, 100).
Use MENU to scroll forward.
Use UNIT to scroll backward.
Fill number of pieces into the container according to the
selected reference quantity.
Wait for the stability display ( ), then confirm by
pressing TARE.
The balance is now in parts counting mode and counts
all units on the weighing plate.
ABS-N_ABJ-NM-BA-e-1614
. Using PRINT the reference weight (UW= Unit weight)
1.0001
] is called up.
59