14.3 Formula mode

The formula function allows to add on various components of a mixture. For check
purpose, the weight of all components (CMP001, (CMP002 etc.), as well as the total
weight (TOTAL) can be printed out. The balance works with a separated memory for
the weight of the weighing container and of the formula components.
In formula mode the „Auto zero" function is not active (see chpt. 12.2).
1. Connect printer (see chpt. 15 „Data output")
2. Call up menu
 In weighing mode press the MENU button twice
3. Select formula mode
 Press the navigation buttons ( ) repeatedly until
„APL.FUNC" is displayed.
 Confirm with PRINT
 Press the navigation buttons ( ) repeatedly
until „FORMULA" is displayed. The formula symbol
[
] appears.
 Press TARE. The indicator
here on the balance is in formula mode.
Enable the edition „component no."(see chpt. 14.3.1)
and „Total" (see chpt. 14.3.2).
4. Weighing components
If necessary, place an empty container on the balance
and tare it by pressing TARE.
 Press PRINT , the symbol
When the GLP function (see chpt. 8.3) is activated, the
headline will be printed.
 Weigh first component.
 Press PRINT.
When standstill control ( ) is completed, the weighing
value of the first component (CMP001) is issued to the
optional printer. The displayed value is added into the
total adding memory. This is followed by the automatic
taring, the symbol [
64
is displayed, from
goes out.
] is displayed.