LED
Color / Status
PP and S Red / Flashing
alternating
PP
Green / Solid
PP
Orange / Solid
S
Orange / Solid
S
Orange / Flashing
S
Green / Flashing
Electrical Conformity
Electromechanical
Compatibility
Environmental Conditions
Ingress of Water
Flammability
Mode of Operation
Dimensions
Weight
Beeps
Condition
3 beeps
Probe or driver failure.
None
Driver primed (probe
retracted).
None
Device is performing
either the prime or pierce
function.
None
Device is engaged
in sample acquisition
sequence.
2 beeps
Sample cup obstructed.
None
Sample cup has been
reinserted (observed after
This medical equipment has passed all required testing for electric shock,
60601-1, CAN/CSA C22.2 No 601-1.
Portable and mobile radio frequency communications equipment (cell
phones) should not be used at close distances. Power frequency magnetic
commercial or hospital environment.
Class II, internally powered Type BF equipment.
Operate the device in ambient conditions (between 15ºC to 25ºC, 30% to
75% relative humidity and 700 to 1060 hPa atmospheric pressure). Trans-
port and store in a cool dry place between -20ºC to +50ºC, 10% to 90%
relative humidity and 500 to 1060 hPa atmospheric pressure.
Not protected against the ingress of water. IPX0.
Continuous.
355g
Action Required
Remove probe. Reset driver by pressing the
PP and S button simultaneously for at least
3 seconds until the motors engage. If reset
red, place driver on charging stand and wait
for driver to reset (PP and S lights will cycle
from red to orange to green to off). If the
indicator lights do not turn off or the probe
cannot be removed, contact Bard Service
and Repair.
Press "PP" when in desired location to
pierce target.
Wait for driver to complete function and
indicator light to turn solid green before
moving or repositioning.
Wait for driver to complete sample acquisition
sequence and indicator light to turn solid
green before moving or repositioning.
Remove sample cup. Inspect cup base and
below cup and remove any tissue outside
green when obstruction is resolved. Press
and hold S button for 3 seconds to reset
remove the biopsy probe, reset the driver,
and replace with a new probe if additional
samples are required.
Press and hold sample button for at least
3 seconds to reset probe. S will turn solid
orange while reset occurs. When reset is
complete both PP and S will be solid green
and the biopsy can be continued.