Note:
integrated coaxial cannula. Then, holding
coaxial cannula in place, remove probe
from breast.
8.
Remove the tissue samples from the sample container by twisting the lid counter clockwise. If
additional samples are required, ensure the sample container lid is in place while acquiring samples.
NOTE: The sample container may be removed completely from the probe by pulling up on the lid
and may be removed before or after the probe is removed from the driver. If the probe is loaded
into the driver and the sample container is removed from the probe, the "PRIME/PIERCE" and
probe, the "PRIME/PIERCE" and "SAMPLE" indicator lights will continuously illuminate green.
9.
Remove the probe from the driver by pressing down on the locking tab, pushing the carriage
completely forward and pulling the probe straight from the driver (Figure 8).
NOTE: Once a probe has been removed from the driver following a procedure, the probe cannot
be re-inserted. If additional tissue samples are required, insert a new probe into the driver.
NOTE: Driver will initiate mechanical reset following probe removal. Do not insert a new probe
prior to completion of this reset. Five beeps and "PRIME/PIERCE" and "SAMPLE" indicators
information.
NOTE: The probe, coaxial cannula and sample container are for single use only. After use, these
products may be a potential biohazard. Handle and dispose of in accordance with accepted
medical practice and applicable local, state and federal laws and regulations.
I. Cleaning and Maintenance
or a damp cloth moistened with a 10% bleach solution. Other than routine cleaning, no other
preventive maintenance is required. However, it is recommended to return the device to an
authorized Bard Service & Repair Facility once per year for inspection and servicing.
the warranty.
will void the warranty.
is inappropriately cleaned it may cause the driver to malfunction and will void the warranty.
Do not autoclave. Do not heat over 54°C (129°F).
The charging stand may be cleaned with same methods as the driver; however, avoid wiping the
charging terminal area. Do not clean charging stand while plugged in.
F
U
®
Biopsy Probe Instructions for Use for additional important
INESSE
LTRA
Figure 7
Figure 8
5
®