The power supply cord cannot be replaced by the user. In case the power supply cord
19
becomes damaged, contact Philips Respironics customer service for replacement (there is
no service option).
Do not over fill the storage compartment with accessories.
20
Before performing any cleaning operation or filter replacement, disconnect the device
21
from the main supply by unplugging it.
Small parts can be inhaled, swallowed or dangerous. Cable, due to its length, may result in
22
provoking strangulation or asphyxiation. Do not leave the device alone with a small child or
physically challenged individual or an individual with learning difficulties.
If you no longer use the device, dispose of according to the local regulations in force.
23
Remember to:
24
• Use this device only with medications prescribed by your doctor;
• Use the accessories recommended by your doctor.
Instructions for use
Remove the compressor from carrying case if supplied before use.
After unpacking the device, check the compressor and accessories for visible damage or
1
defects such as cracks in the plastic housing which could expose electrical components.
Contact Philips Respironics customer service or your product distributor to report any
damaged product.
When taking a treatment, sit upright and relaxed.
2
Plug the compressor into the electrical outlet.
3
Refer to the nebulizer and accessory instructions for use before treatment.
4
To start the treatment, switch the device ON by pressing the ON / OFF switch onto
5
the "I" position.
When the treatment has been completed, switch the unit OFF by setting the switch
6
position to "O" and disconnect the plug from the electrical outlet.
Do not close the lid when the nebulizer tubing is attached.
7

Cleaning and maintenance

During cleaning, make sure the power cord is not plugged in the electrical outlet.
Cleaning of the compressor, nebulizer and accessories
Use a damp cloth to wipe down the outside of the compressor at least once each month.
Refer to nebulizer and accessory instructions for use for cleaning of the accessories and nebulizer.
6
Replacement of the nebulizer
The nebulizer and other accessories should be replaced according to its instructions.
Replacement of the filter
The filter is to be replaced when it appears discolored or wet.
Troubleshooting
The device does not turn on
• Make sure the plug is firmly fitted to the electrical outlet.
• The thermal protector may have been activated by the following reasons:
a) the device was used near a heater or in a hot environment
b) the air vent openings (6) are obstructed and do not allow for the motor ventilation
If this is the case set the ON/OFF switch to 'O' and let is sit for at least 60 minutes. In the
event of malfunction (insufficient pressure and flow) turn the unit off for 10 seconds and
turn back on.
After each cleaning, check that the unit is working properly by verifying that air still comes
out of the nozzle when the unit is turned to the "ON" position.
Unauthorized repair voids warranty.
If the device does not work properly after attempting these troubleshooting steps, contact your
authorized product distributor or Philips Respironics Customer Service at +1 724 387 4000.
Maintenance and repairs
Never open the device. There are no serviceable parts within the unit. The compressor
requires no lubrication or maintenance.
Technical specifications
Electricity supply
230V/50Hz. 1 Amp
Thermal overload protector
Resettable thermal fuse; functioning temperature 150 C°
Compressor max. pressure
360 kPa (3.6 bar)
Average flow rate
7 LPM @ 70 kPa (0.7 bar)
Max flow rate
9.8 LPM
Weight
2.2kg
Size
380mm x 170mm x 110mm
Noise level
58±3dBA
ENGLISH
7