Philips RESPIRONICS InnoSpire Deluxe Mode D'emploi page 97

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 28
A. Instructions for use
1. Innospire Deluxe needs special precautions regarding EMC and needs to be installed and put into service
according to the EMC information provided in this document;
2. Portable and mobile RF communications equipment can affect Innospire Deluxe.
B. Technical description
1. Warning that the use of accessories, transducers and cables other than those specified with the exception
of transducers and cables sold by the manufacturer of the Innospire Deluxe as replacement parts for inter-
nal components, may result in increased EMISSIONS or decreased IMMUNITY of the Innospire Deluxe.
2. Warning that the Innospire Deluxe should not be used adjacent to or stacked with other equipment.
3.
Guidance and manufacturer's declaration – electromagnetic emissions
The Innospire Deluxe is intended for use in the electromagnetic environment specified below. The customer or the user
of the Innospire Deluxe should assure that it is used in such an environment.
Emissions
Compliance
RF emissions
Group 1
CISPR 11
RF emissions
Class B
CISPR 11
Harmonic emissions
Class A (not applicable for 120v
rated devices)
IEC 61000-3-2
Voltage fluctuations/flicker emissions
Complies (applicable to 50Hz
rated devices only)
IEC61000-3-3
190
Electromagnetic environment – guidance
The Innospire Deluxe uses RF
energy only for its internal function.
Therefore, its RF emissions are
very low and are not likely to
cause any interference in nearby
electronic equipment.
The Innospire Deluxe is suitable for
use in all establishments, including
domestic establishments and those
directly connected to the public
low-voltage power supply network
that supplies buildings used for
domestic purposes.
4.
Guidance and manufacturer's declaration – electromagnetic immunity
The Innospire Deluxe is intended for use in the electromagnetic environment specified below. The customer or the user
of the Innospire Deluxe should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electrostatic discharge
±6 kV contact
±6 kV contact
(ESD)
±8 kV air
±8 kV air
IEC 61000-4-2
Electrical fast
±2 kV for power
±2 kV for power
transient/burst
supply lines and
supply lines and
dc lines
dc lines
IEC 61000-4-4
Surge
±1 kV line(s)
±1 kV line(s) and
and neutral
neutral
IEC 61000-4-5
±2 kV line(s)
±2 kV line(s)
to earth
to earth
Voltage dips, short
<5 % U
<5 % U
T
T
interruptions and
(>95 % dip in U
)
(>95 % dip in U
T
voltage variations
for 0,5 cycle
for 0,5 cycle
on power supply
40 % U
40 % U
input lines
T
T
(60 % dip in U
)
(60 % dip in U
T
IEC61000-4-11
for 5 cycles
for 5 cycles
70 % U
70 % U
T
T
(30% dip in U
)
(30% dip in U
T
for 25 cycles
for 25 cycles
<5 % U
<5 % U
T
T
(>95 % dip in U
)
(>95 % dip in U
T
for 5s
for 5s
Power frequency
3 A/m
3 A/m
(50/60Hz) magnetic
field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
ENGLISH
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be
at least 30%.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
)
typical commercial or hospital environment.
T
If a dips or an interruption of mains power
occurs, the current of the Innospire Deluxe
may be dropped off from normal level, it may
)
be necessary to use uninterruptible power
T
supply or a battery.
)
T
)
T
Power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.
191

Publicité

Table des Matières
loading

Table des Matières