Guidance and manufacturer's declaration – electromagnetic immunity
The Innospire Deluxe is intended for use in the electromagnetic environment specified below. The customer or the user
of the Innospire Deluxe should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Conducted RF
3Vrms
3Vrms
IEC 61000-4-6
150kHz to 80MHz
Radiated RF
3V/m
3V/m
IEC 61000-4-3
80MHz to 2.5GHz
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection
from structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the Innospire Deluxe is used exceeds the applicable
RF compliance level above, the Innospire Deluxe should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating the Innospire Deluxe.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
192
Electromagnetic environment – guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the Innospire Deluxe, including cables,
than the recommended separation distance
calculated from the equation applicable to the
frequency of the transmitter.
Recommended separation distance
d = 1.2√P
d=1.2√P 80MHz to 800MHz
d=2.3√P 800MHz to 2,5GHz where P is
the maximum output power rating of the
transmitter in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
metres (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
a
survey
, should be less than the compliance
level in each frequency range
b
. Interference
may occur in the vicinity of equipment
marked with the following symbol:
6.
Recommended separation distances between portable and mobile RF communications equipment
and the Innospire Deluxe
The Innospire Deluxe is intended for use in an electromagnetic environment in which radiated RF disturbances are
controlled. The customer or the user of the Innospire Deluxe can help prevent electromagnetic interference by maintaining
a minimum distance between portable and mobile RF communications equipment (transmitters) and the Innospire Deluxe
as recommended below, according to the maximum output power of the communications equipment.
Separation distance according to frequency of transmitter
Rated maximum
output power of
transmitter
150 kHz to 80 MHz
W
d=1.2√P
0.01
0.12
0,.1
0.38
1
1.2
10
3.8
100
12
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in metres
(m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
ENGLISH
m
80 MHz to 800 MHz
800 MHz to 2,5 GHz
d=1.2√P
d=2.3√P
0.12
0.23
0.38
0.73
1.2
2.3
3.8
7.3
12
23
193