F Quick Release Tabs
G Tubing Assembly
H Swivel
I Elbow with Built-in Exhalation (do not block)
J Frame Top Slot
K Headgear Top Strap
L Headgear Crown Strap with Loop for Tube Management (optional)
M Headgear Back Strap
N Headgear Bottom Strap
Before Use Read and Understand the Instructions Completely.
• Hand wash the entire mask a.
• Wash face. Do not use moisturizer/lotion on hands or face.
• Inspect the mask. Replace the cushion if hardened or torn. Replace any parts that are broken or
worn.
• If the mask has magnetic headgear clips, ensure that they are assembled properly into the frame
and headgear.
Cleaning Instructions
Note: If your mask has magnetic headgear clips, leave them attached to the frame before
cleaning.
Daily: Hand wash the non-fabric parts.
Weekly: Hand wash the fabric parts.
1. Wash in warm water with liquid dishwashing detergent.
2. Rinse thoroughly.
3. Air dry completely before use.
Cautions:
Do not clean the mask with bleach, alcohol, solutions containing bleach or alcohol, or solutions
•
containing conditioners or moisturizers.
Inspect the mask for damage or wear (cracking, crazing, tears, etc). Discard and replace any
•
components as necessary.
Any deviation from these instructions will impact the performance of the product.
•
Institutional Disinfection
Note: If your mask has magnetic headgear clips, remove them from the frame before institutional
disinfection. Refer to the Disassembly/Assembly of Magnetic Headgear Clips section for removal.
Refer to the Disinfection Guide for Professional Users to reprocess between patients in a clinical
setting. Access the latest version of the Disinfection Guide at www.healthcare.philips.com or by
contacting customer service at 1-800-345-6443 (USA or Canada only) or +1-724-387-4000.