Philips RESPIRONICS Wisp Nasal Mask Mode D'emploi page 7

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

Specifications
Warning: The technical specifications of the mask are provided for the healthcare professional
to determine if it is compatible with the CPAP or bi-level therapy device. If used outside these
specifications, or if used with incompatible devices, the mask may be uncomfortable, the seal of the
mask may not be effective, optimum therapy may not be achieved, and leak, or variation in the rate
of leak, may affect device function. The pressure flow curve shown below is an approximation of
expected performance. Exact measurements may vary.
Pressure Flow Curve
60
50
40
30
20
10
0
0
Resistance
Drop in Pressure at
All sizes
Deadspace Volume
P
S/M
L
XL (adult only)
Sound Levels
A-weighted Sound Power Level 27 dBA
A-weighted Sound Pressure Level at 1 m 19 dBA
Disposal
Dispose of in accordance with local regulations.
Storage Conditions
Temperature: -4º to 140º F (-20º to +60º C)
Relative Humidity: 15% to 95% non-condensing
5
10
Pression cm H
50 SLPM
0.8 cm H
O
2
37 mL
49 mL
73 mL
85 mL
15
20
O
2
100 SLPM
3.2 cm H
O
2
7
25
30

Publicité