Les langues disponibles

Filter
AC Filter
Baseline Filter
EMG Filter
Low Pass Filter
5.5.8 Auto-Analysis
Auto- Analysis
Auto Analysis
Diagnostic
Classification
5.5.9 Others
Other settings include demo mode, 6-leads mode, lead off alarm, heart beat sound, pacemaker
sensitivity and password modification, as shown in the following table:
Three options of 50Hz, 60Hz and off are provided. The purpose of AC filter opening is
to eliminate the interference of AC power supply. The selection of AC filter value shall
be set according to the natural frequency of national AC power supply.
There are 7 options: 0.05 Hz, 0.16 Hz, 0.25 Hz, 0.32 Hz, 0.5 Hz, 0.67 Hz and off. Baseline
filter makes the baseline of ECG waveform regress rapidly, and the larger the filtering
value, the faster the baseline regression.
There are 7 options: 20 Hz, 25 Hz, 30 Hz, 35 Hz, 40 Hz, 45 Hz and off. The smaller the
filtering value is, the stronger the ability of eliminating EMG interference is.
There are four options: 75 Hz, 100 Hz, 150 Hz and off.
Enable this item, the measurement parameters and analysis interpretation can be
automatically generated after more than or equal to 10 seconds' acquisition.
Enable this item, ECG at the end of acquisition more than or equal to 10s will be
classified according to the analysis interpretation. The software includes normal ECG,
borderline ECG, atypical ECG and abnormal ECG.
www.spengler.fr
Function Introduction
Function introduction
105