2. Select the printing format through the 【Print】 drop-down triangle, and then print the ECG
report according to the selected printing format.
3. Right click the replay interface and select 【Print】 to print the ECG report in the default
print format.
4.3.3.2 Auto Print
«System settings» → «print settings» to select automatically print the report. ECG acquisition
and analysis software for Windows PC will automatically print the ECG report according to the
set conditions after the acquisition stops.
4.3.3.3 Print Preview
The current ECG report can be
previewed when printing the report.
«System settings» → «Print setting» to
choose print preview function when
printing». When printing is selected,
enter the print preview interface first.
4.4 Reanalysis
In the replay interface of ECG waveform, click 【Reanalysis】 to enter the interface of
reanalysis, as shown in the figure below. The interface of reanalysis shows ECG characteristics,
average template of each lead, interpretation, etc., and ECG parameters can be re-analyzed.
Modify parameters: users can manually input the parameters or move the location line of the
average template to modify the parameters.
Switch display channel: users can click the lead name to switch to display the average template
of the corresponding lead.
Modify conclusion:
 Add: directly input the conclusion in the conclusion input box manually or select the
conclusion in the analysis result index, and click [Add] to add the conclusion.