2. Start Sampling
In the patient information interface, click 【OK】 to start sampling ECG data.
Presampling: the «Presampling Time» of the "System Settings"- >» Device and Sample
Settings «is set to the non close state. ECG data under presampling mode will not be
recorded.
After the set presampling time, the system starts to record ECG data, and the recording time
of the main interface starts to count.
In the acquisition process, click 【Sampling】 to clear the recording time and restart
recording ECG data for the current patient.
It is strongly recommended that when there is strong interference in the current acquisition
process, click 【Sampling】 again to eliminate the interference after the interference
disappears, so as to obtain high-quality ECG waveform and reduce misdiagnosis.
3. Stop
Manual stop: click 【Stop】 to end the current acquisition. When the recording time is less
than 10s, the software does not save the data.
Auto stop: if Auto is selected in "Settings" - > "Device and Sample Settings ", "Sample
Mode", ECG acquisition and analysis software for Windows PC will automatically stop and
save the file after the record reaches the set sampling time.
If the waveform is replayed after the acquisition is stopped, the quality of the waveform
is poor. It is strongly recommended to re-sample the patient to obtain high-quality ECG
waveform and reduce misdiagnosis.
4.3.2 Check the Quality of ECG
4.3.2.1 Lead off Alarm
During the acquisition process, if the lead connection is unstable or dropped, the software
will prompt the alarm of lead off. Check the loose electrode or lead according to the prompt
information, and reconnect the electrode or lead.
Note:
1) When there is a lead off alarm during the acquisition process, it is suggested to acquire
again.
2) The data acquisition under the condition of lead off may lead to information loss, missed
diagnosis or misdiagnosis.
4.3.2.2 Eliminating Waveform Interference
If there is interference in the waveform, please check the filter settings so that the filter settings
are most suitable for the current environment.
If the filter cannot eliminate the interference, please check the electrode connection and skin
condition of the patient, and check whether the ambient temperature is appropriate.
If the above steps still can not eliminate interference, please check whether there is strong
interference equipment in the surrounding environment.
4.3.3 Report print
4.3.3.1 Manually print
Manual printing can be realized in the following three ways:
1. Click 【Print】 on the toolbar to print the ECG report in the default print format.
Save the current input patient information and enter the main interface to start
sampling.
Do not save the current input information and return to the main interface.