4.6 System Settings
Click 【Settings】 to open the administrator password input dialog box. The administrator
password is 123456 by default. Enter the password and click 【OK】 to enter the system setting
interface. The user can set the hospital information, device and sample settings, rhythm lead,
print, export, display format, display color, filter, FTP (optional), DICOM (optional), Worklist
(optional) , HL7 (optional), upload (optional), path, other relevant information according to the
actual needs.
Name
Hospital information
settings
Device & Sample
Settings
Import ECG data from disk, support EM-XML, FDA-XML, SCP, BKG, DICOM standard
format ECG data.
Support the export of ECG data from software to disk.
Select one or more patient records and click 【Upload】to upload the selected
patient record data to the corresponding server according to the set upload method.
Upload status:
Click 【Close】 or click the close icon «X» in the upper right corner to exit the
patient record management interface and return to the main collection interface.
Including hospital title, user-defined term and auto fill .
1) The Device, COM NO., Device NO., Sample Mode, PreSampling Time and Sensitivity
of Pace Detection can be set.
2) Device: USB ECG device and Bluetooth ECG device. COM No.: select the type of
Bluetooth outgoing COM port; only set when the "device" is Bluetooth ECG Device.
3) Device NO.: represents the unique number of the equipment, and can be input in
English, numbers and special symbols.
4) Sampling Time in Auto Mode: it can be entered manually to control the sampling
time in automatic record mode; this setting is only for automatic record mode.
5) Sample Mode: manual and auto
6) PreSampling Time: 3 seconds, 5 seconds and 10 seconds can be selected. It is off by
default.
7) Sensitivity of Pace Detection: including low, medium, high and off. When the patient
has implanted pacemaker, it is necessary to turn on the pacemaker detection sensitivi-
ty function, otherwise the pacemaker signal cannot be detected.
www.spengler.fr
Function
indicates uploaded
Function
91
indicates not uploaded.