Télécharger Imprimer la page

Spengler CARDIOMATE PC Manuel Utilisateur page 122

Publicité

Les langues disponibles

Les langues disponibles

Table 2
Guidance and manufacturer's declaration - electromagnetic Immunity
Immunity Test
Electrostatic discharge (ESD)
IEC 61000-4-2
Electrical fast transient/burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short interruptions
and voltage variations
on power supply input lines
IEC 61000-4-11
Power frequency magnetic field
IEC 61000-4-8
Conduced RF
IEC61000-4-6
Radiated RF
IEC61000-4-3
NOTE: UT is the a.c. mians voltage prior to application of the test level.
Table 3
Guidance and manufacturer's declaration - electromagnetic Immunity
(Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment)
Test
Band
Frequency
(MHz)
(MHz)
385
380 –390
450
380 –390
710
745
704 – 787
780
810
870
800 – 960
930
IEC 60601-1-2 Test level
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Power supply lines:±2 kV
input/output lines: ±1 kV
Line(s) to line(s) : ±1 kV.
Line(s) to earth : ±2 kV.
100 kHz repetition frequency
At 0º, 45 º, 90 º, 135 º, 180 º, 225 º,
70% 25/30 cycles
Single phase: at 0
150KHz to 80MHz: 3Vrms
6Vrms (in ISM and amateur radio
Radiated RF IEC61000-4-3
Service
Modulation
Pulse modulation
TETRA 400
GMRS 460,
± 5 kHz deviation
FRS 460
1 kHz sine
Bande LTE
Pulse modulation
13, 17
217 Hz
GSM 800/900,
TETRA 800,
Pulse modulation
iDEN 820,
CDMA 850,
Bande LTE 5
www.spengler.fr
±8 kV Contact
0% 0,5 cycle
270 º et 315 º
0% 1 cycle
and
0% 300 cycle
30 A/m
50Hz/60Hz
bands)
80% Am at 1kHz
10 V/m
80MHz– 2,7GHz
80% AM to 1kHz
Modulation
(W)
1.8
18 Hz
FM
2
0.2
2
18 Hz
122
Compliance level
±8 kV Contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Power supply lines:±2 kV
Line(s) to line(s) : ±1 kV.
100 kHz repetition frequency
0% 0,5 cycle
At 0º, 45 º, 90 º, 135 º, 180 º, 225 º,
270 º et 315 º
0% 1 cycle
and
70% 25/30 cycles
Single phase: at 0
0% 300 cycle
30 A/m
50Hz/60Hz
150 KHz to 80 MHz : 3Vrms
6Vrms (in ISM bands)
80% Am à 1kHz
10 V/m
80MHz– 2,7GHz
80% AM to 1kHz
IMMUNITY
Distance
TEST LEVEL
(m)
(V/m)
0.3
27
0.3
28
0.3
0.3
28
9

Publicité