Spengler CARDIOMATE PC Manuel Utilisateur page 96

Publicité

Les langues disponibles

Les langues disponibles

Icon
Button Name
New patient
Sampling
Stop
Paper speed
Gain
Filter
Setting
Data base
Exit
5.3 ECG Acquisition and Report Print
The ECG acquisition and analysis software for Android Tablet can acquire ECG data through
USB wired acquisition and Bluetooth wireless acquisition. Before sampling ECG data, please
make sure that the patient has connected the electrode, the patient cable correctly connected
to the ECG acquisition box, and the acquisition box reliably connected to the tablet. See [3.Be-
fore Operation] for details.
Note :
1) In case of waveform pause or acquisition interruption during acquisition, please check the
reliability of connection between acquisition box and software.
2) When the waveform interference is large or the lead falls off during the acquisition, please
check the reliability of the connection between the patient cable and the acquisition box
and the human body.
3) In the acquisition process, if the distance between software and Bluetooth ECG acqui-
sition box exceed the effective transmission distance, the acquisition will be interrupted
automatically.
5.3.1 Input Patient Information
Open the ECG acquisition and analysis software for Android Tablet, click 【New Patient】on
the toolbar, enter the corresponding information in the pop-up "Patient Information" dialog
box, and click 【OK】:
Click 【New patient】 to open the patient information dialog box, enter the
new patient information, and click 【OK【 to start sampling ECG data for the new
patient.
Click 【Sampling】 to acquire ECG data directly.
Click 【Stop】to stop ECG data acquisition.
Click 【Paper speed】 to adjust the paper speed of waveform.
Click 【Gain】 to adjust the waveform gain.
Note: when the gain is set to 10 / 5 mm / mV, 20 / 10 mm / mV and AGC
(automatic gain), the gain of the main interface is displayed as 10 mm / mV.
If the gain of the channel is not 10 mm / mV, the multiple relationship is added
next to the lead symbol, where (x2), (x4), (x1 / 2), (x1 / 4), (x1 / 8) are
respectively represented as the gain of the current channel as 20 mm / mV,
40 mm / mV, 5 mm / mV, 2.5 mm / mV and 1.25 mm/mV.
Click 【Filter】 to adjust the current ECG filter parameter value during
acquisition.
Filter includes AC filter, EMG filter, baseline filter and low pass filter.
Click 【Setting】 to enter the system setting interface, where you can set related
settings.
Enter the database, you can query, edit, export, upload (optional) and other
operations for the record patient record.
Exit the system.