Cartridge Replacement and Calibration

Changing the Sensor Cartridge

Cartridge Replacement and Calibration
Changing the Sensor Cartridge
Since the sensor module is completely replaceable, you can simply
open the 301W transmitter, remove the depleted or defective S301W
cartridge and insert a new S301W cartridge. The sensor module is
equipped with a 16 pin (female) connector and needs only to be
pushed onto the connector pins in the 301W transmitter.
If the removed sensor is defective, return to Honeywell Analytics
according to the procedures described in the Warranty section of the
manual. If the unit has simply reached its End-of-life, discard according
to local Hazardous-Materials regulations (pcb and battery disposal).
Never remove the S301W cartridge when the transmitter is
functioning. Always power-off before opening the transmitter
CAUTION
(see How to Program the Transmitter; Function 9 - Save
Parameters and Close Device).