Instrument Status Indicators
Status
ON
Fault:
Self-test fail
Life ending
warning
Life ended
Low alarm
High alarm
Over range
alarm
Power off
Alarm Output Setpoints
Gas alarm setpoints at Alarm output 1 and alarm output 2 activate field
relays, alarms etc. Special setpoints are available.
Order Number
TP1-M
When the gas levels return to acceptable ranges, the gas alarm stops.
511407
6/07
LED Indicator
ON
Fast flashes; 1
every 0.5
seconds
Slow flashes; 1
every 2 seconds
Off
On
On
On
OFF
Low Alarm Setpoint
25 ppm
ToxyPoint User Manual
Honeywell
Output to
Alarm Output Signal
Controller
4-20 mA
Triggers low alarm O/P1
2 mA
and high alarm O/P2
4-20 mA
2 mA
4-20 mA
Triggers O/P1
4-20 mA
Triggers O/P1 and O/P2
24 mA
Triggers O/P1 and O/P2
0 mA
High Alarm Setpoint
200 ppm
Operation
Using the Sensor
21