Invacare Reliant Plus RHL450-1 Instructions D'installation Et D'utilisation page 25

Table des Matières

Publicité

Les langues disponibles

Les langues disponibles

CARE AND MAINTENANCE OF YOUR PATIENT LIFT
Front Casters
1. Place the stand up lift on its side.
2. Remove the bolt and locknut that secure the exist-
ing front caster assembly to the fork.
Base
Locknut
Bolt
Rear
Caster
Bushing
Locknut
Washer
Front
FRONT
Casters
FIGURE 5 - REPLACING CASTERS AND/OR
FORKS
CARE AND MAINTENANCE OF YOUR PATIENT LIFT
REAR
Fork
Base
Fork
Bolt
NOTE: The front caster assembly consists of two (2)
casters and a washer in between.
NOTE: The washer will fall out from between the two
(2) casters.
NOTE: Existing washer will be reused. Examine and
replace if worn.
3. Position new/existing washer between the two (2)
new casters.
4. Line up the mounting holes in the new front caster
assembly and the fork.
5. Install the bolt through the fork and new front caster
assembly and tighten securely with the locknut.
Forks
1. Place the stand up lift on its side.
2. Remove the front or rear caster from the stand up
lift. Refer to REPLACING CASTERS in this sec-
tion of the manual.
3. Unscrew the existing fork from the base.
4. Install the new fork onto the base.
5. Install the front or rear caster onto the stand up lift.
Refer to REPLACING CASTERS in this section
of the manual.