For more information about the residence time and concentration
of disinfectants, please contact your disinfectant dealer or the
manufacturer of the disinfectant.
7.4 Checking and Tightening Mast Pivot Bolt
A
B
Check that the bolt A is through the bracket B and the locknut
1.
C is tight and secure.
2.
If needed, do one or more of the following:
•
Tighten locknut and back-off the locknut 1/8 of a turn.
•
Replace the locknut.
7.5 Checking the carabiner and its mounting
WARNING!
Risk of Injury
Worn or damaged parts of the lift may cause injury to
the patient or assistants.
– After the first year of use, the hooks of the hanger
bar and mounting brackets of the boom should
be inspected every six months to determine the
extent of wear. Check for signs of cracking, fraying,
deformation or deterioration. If these parts become
worn, replacement must be made.
38
C
Checking the carabiner
Check the carabiner for wear at contact points A and B.
1.
Do not use the lift if the carabiner measures less than 6 mm at
these points.
Checking the boom pin
Remove plastic part A.
1.
Turn carabiner B up to see the welded pin C.
2.
Check the pin C for wear.
3.
Do not use the lift if the pin measures less than 7 mm.
4.
Insert the plastic part again.
7.6 Replacing the Hanger Bar
If the scale is installed on the lift, refer to the scale instruction
manual to replace the hanger bar.
A
B
C
A
B