Surge
± 1 kV line(s) to line(s)
IEC 61000-4-5
< 5% U
for 0,5 cycle
Voltage dips, short
interruptions and
40% U
voltage variations on
5 cycles
power supply input
70% U
lines
25 cycles
IEC 61000-4-11
< 5% U
for 5 sec
Power frequency
(50/60 Hz) magnetic
3 A/m
field
IEC 61000-4-8
1575480-B
± 1 kV line(s) to line(s)
(>95% dip in U
)
< 5% U
T
T
for 0,5 cycle
(60% dip in U
) for
40% U
T
T
5 cycles
(30% dip in U
) for
70% U
T
T
25 cycles
(>95% dip in U
)
< 5% U
T
T
for 5 sec
30 A/m
Mains power quality should be that of a typical commercial
or hospital environment.
Product is double-insulated. There are no other possible
connections to earth
(>95% dip in U
)
T
T
Mains power quality should be that of a typical commercial
or hospital environment. If the user of the patient
(60% dip in U
)for
lift requires continued operation during power mains
T
T
interruptions, it is recommended that the patient lift be
powered from an un-interruptible power supply or a
(30% dip in U
) for
T
T
battery.
U
is the a. c. mains voltage prior to application of the
T
(>95% dip in U
)
T
T
test level.
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial
or hospital environment.
Technical data
47