4.12 Change the analyzer or channel names
1. Push menu, then select SETUP SYSTEM > CONFIGURE ANALYZER.
2. Select an option.
Option
EDIT ANALYZER
NAME
EDIT CHANNEL
NAME
4.13 Start or stop measurements on a channel (2- or 4-channel analyzers)
1. Push menu, then select SETUP SYSTEM > CONFIGURE ANALYZER > CONFIGURE
SEQUENCER > ACTIVATE CHANNELS.
2. Select a channel to start measurements. Unselect a channel to stop measurements. Push the
LEFT arrow to select or unselect a checkbox.
4.14 Change the channel measurement order (2- or 4-channel analyzers)
To change the order that the channels are measured, do the steps that follow:
1. Push menu, then select SETUP SYSTEM > CONFIGURE ANALYZER > CONFIGURE
SEQUENCER > SEQUENCE CHANNELS.
2. Push the UP and DOWN arrows to select a row.
Note: S1 is the first channel measured, followed by S2, S3 and S4.
3. Push the LEFT or RIGHT arrow to select a channel.
Note: Do not select channels that contain the symbol "~" (e.g., 4-~SAMPLE4). Channels that contain the
symbol "~" are not measured.

4.15 Set the date and time

Set the date and time format and the date and time that show on the measurement screen and in the
log files.
1. Push menu, then select SETUP SYSTEM > CONFIGURE ANALYZER > SET DATE/TIME.
2. Select an option.
Option
DATE FORMAT Sets the date format (YYYY= year, MM=month and DD=day) and time format
DATE/TIME
4.16 Configure the 4-20 mA analog outputs
If an analog output(s) in the analyzer is connected to an external device, select the channel
represented at the analog output and the measurement range.
1. Enable an analog output as follows:
a. Push menu, then select SETUP SYSTEM > CONFIGURE ANALYZER > SETUP OUTPUTS
> 4-20 mA SETUP > [select an output].
b. Select SELECT SOURCE > [analyzer name].
Description
Changes the name of the analyzer. Enter a unique name, such as the
analyzer location (16 characters maximum). The analyzer name shows
on the measurement screen and the data logs.
Changes the name of the selected channel. Enter a unique name, such
as the source of the sample water (10 characters maximum). The
channel name(s) shows on the measurement screen and the data logs.
Description
(12-hour or 24-hour). Default: YYYY-MM-DD 24 hours.
Sets the date and time. Use the arrow buttons to enter the date and time.
English 13