Set The Date And Time - Hach A1000 Manuel D'utilisation De Base

Masquer les pouces Voir aussi pour A1000:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 29
Flush the analyzer
The cleanliness of commercially available tubing is not consistent
(particularly stainless steel). The manufacturer recommends that a clean
cycle is done at initial startup to fully flush the sample inlet tubing. Refer
to
Do a clean cycle
on page 24.
Start automatic analysis
Put the analyzer in Auto TOC mode. Refer to
on page 24.
The analyzer is calibrated at the factory. For accurate results, let the
analyzer complete five analyses before the reported data is accepted as
correct.
If at any time the sample is not within the analyzer specifications, an
alarm occurs (e.g., "Code 38" or "Code 40"). Refer to and .
Give the analyzer a channel ID
Each analyzer must have a unique channel ID (1–8) to identify the
analyzer. The channel ID is the network address for the analyzer and
corresponds to a channel LED on the controller. The channel ID is
shown before the analyzer name on the controller display. Duplicate
channel IDs are not permitted on the network.
Note: The controller sets the default channel ID for each analyzer when it first
scans it.
1. At the controller, select the analyzer, then push SETUP.
2. Use the UP and DOWN arrows to select System Setup>ID to SN
Xref. The Channel ID–Name is shown for each analyzer connected
to the controller. The first Channel ID is highlighted.
3. Use the UP and DOWN arrows to change the channel ID. Push
ENTER to move the cursor to the next channel ID.
Note: Duplicate channel IDs can be entered while in the edit mode, but they
cannot be saved.
4. When no more changes are necessary, push ESC.
Note: If the controller identifies duplicate channel IDs, the display shows a
message that identifies the condition and that the conflict must be corrected
before the screen can be exited.
Select the operating mode
5. Push ENTER. The controller resets the network. The main screen is
shown.
Give the controller a network address
Each controller must have a unique network address for identification
and reporting functions (9–16). The default network address of the
controller is 9. It may be necessary to change the network address of the
controller to prevent network conflict.
To change the network address of the controller:
1. At the controller, push down and hold SETUP. Disconnect the power
adapter from the controller, then connect it again. (On an
A1000 S20P or A1000XP, turn the analyzer off and back on.) The
controller display shows the current network (A-Net) address for the
controller.
2. Use the UP and DOWN arrows to select the controller network
address (9–16).
3. Push ENTER to save the change.

Set the date and time

Set the date and time on each controller. The controller transmits the
date and time to the analyzers connected to it so that report functions
are in sync.
1. At the controller, push SETUP.
2. Use the UP and DOWN arrows to select System Setup>System
Time.
3. Push ENTER. The flashing cursor becomes an underscore.
4. Use the UP and DOWN arrows to set the date (dd/mm/yy) and then
the time (hh:mm) in 24-hour format. Push ENTER to move the cursor
to the next interval (e.g., the day or year).
Note: Although only the right digit of each interval is highlighted, all of the
interval is selected.
5. Push ESC multiple times to go back to the main screen.
English 23

Publicité

Table des Matières
loading

Ce manuel est également adapté pour:

A1000xp

Table des Matières