Repairs; Electrical Safety Testing Information; Fault Codes; Table 10. Fault Codes - Johnson & Johnson DePuy Synthes FMS VUE II Manuel De L'opérateur

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 293
9.2

repairs

For repairs, contact Customer Service Center in the USA at +1-800-382-4682 . For repairs
outside the United States, contact your local affiliate . Relevant service instructions are provided
in DePuy Synthes work instructions . DePuy Synthes will provide circuit diagrams, component
part lists, descriptions, calibration instructions, or other information that will assist service
personnel trained and authorized by the manufacturer in repairing the pump .
9.3

electrical safety testing information

For electrical safety testing information, contact Customer Service Center in the USA at
+1-800-382-4682 . Outside the United States, contact your local affiliate .
9.4

fault Codes

The FMS VUE II software detects the following fault conditions .
fault Code
FC001
System Error
(Unrecoverable)
FC002
System Error
(Recoverable)
FC003
Pressure tubing system problem
detected

Table 10. Fault Codes

fault description
56

fault Correction

Discontinue use of the console . Contact
your DePuy Synthes representative .
Power down the pump .
Wait 1 minute before power up .
If fault occurs again, discontinue use of
the console .
Contact your DePuy Synthes
representative .
Pressure sensors readings are matching
but incorrect .
Change the Tube Set . If changing Tube
Set does not correct issue, discontinue
use of the console . Contact your
DePuy Synthes representative .