Operating precautions and general safety

read and understand this manual before using the fms vue ii system.
the following list of warnings and precautions applies to the general
operation of the fms vue ii system.
2.1

warnings

nOte: Tips for optimal operation and other useful information .
CautiOn! Failure to observe caution can result in damage to the device .
warning! Failure to observe warning can result in injury to the patient or user .

2.1.1 electrical safety

• Electrical safety testing should be performed by a biomedical engineer or other qualified
person .
• Risk of electrical shock: Do not remove the pump housing . Refer servicing to authorized Repair
Service Center .
• Do not connect the device to a power source that is not properly earthed (grounded) . If the
power source is not grounded, use the equipotential connector .
• Disconnect the device from the main power source when cleaning, servicing or inspecting .
• Inspect all equipment and cables periodically for wear . If damage is noted, replace or return to
Repair Service Center .
• Avoid fluid contact with the system and its electrical connectors .
• To avoid risk of fire, replace fuses with same type and rating .
• Do not restrict access to the power switch . Ensure that the pump is positioned so that access
to power main is not obstructed . To remove all mains power from the unit, disconnect the
power cord from the appliance inlet .
• The FMS VUE II pump power supply detachable cord serves as the means of isolation
between the power supply and the power main . Ensure that the placement of the equipment
does not interfere with access to the power supply cord .