filling the Chamber

warning! Over-filling the chamber may wet the pressure sensor filter, which
may interfere with accurate pressure measurement. the fill Chamber mark
indicates correct fluid level.
6.4

sOlO mode (irrigation and handpiece Control)

CautiOn! pump tubing must be configured according to instructions in section 5.6.
CautiOn! when using wall suction as the flow source for the shaver, a flow limiter is
required to prevent excessive flow rates, which could produce low (potentially vacuum
level) joint pressures.
to activate sOlO mode:
1 . Press run/stOp to place the pump in STOP mode .
2 . Press the sOlO pad to activate SOLO Mode .
3 . Press the sOlO pad again to deactivate SOLO Mode .
to save and
) to exit
To fill the chamber:
Press the fill Chamber pad to fill the pressure chamber initially, and
throughout surgery as needed . The chamber should never be over
1/3 full .
nOte: If pressure alarm sounds, refer to Section 6.5, Pressure
Warnings and Alarms.