Connecting The Handpiece - Sterile Zone; Connecting The Cannula - Sterile Zone; Tubing Connections For Solo Mode; End Of Operation - Johnson & Johnson DePuy Synthes FMS VUE II Manuel De L'opérateur

Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

  • FRANÇAIS, page 293
1 0 , 0 0 0 m l
9,0 0 0 m l
8,0 0 0 m l
7,0 0 0 m l
6,0 0 0 m l
5,0 0 0 m l
4,0 0 0 m l
3,0 0 0 m l
2,0 0 0 m l
1,0 0 0 m l
5 0 0 m l
5.4
Connecting the handpiece – sterile ZOne
sterile operator: Use this procedure to connect the handpiece in the sterile zone .
Use aseptic technique . Insert the tube labelled "SHAVER" onto the handpiece suction port .
5.5
Connecting the Cannula – sterile ZOne
sterile operator: Use this procedure to connect the Cannula in the sterile zone .
Use aseptic technique . Insert the tube labelled "CANNULA" onto the Cannula suction port .
5.6

tubing Connections for sOlO mode

To operate the pump in SOLO Mode (irrigation and handpiece control only):
before operation:
1 . Make sure the pump tubing is connected as described in in this chapter .
2 . Connect the Inflow Tubing as described in Section 5 .1 .
3 . Connect the Intermediary Tubing as described in Section 5 .2 .
4 . Refer to Section 6 .4 for SOLO Mode Operating instructions .
5.7

end of Operation

at the end Of the OperatiOn, remOve and disCard all tubing.
nOte: The Inflow tubing (Ref . 284508) may be re-used during the same surgical day ONLY
when used with the Intermediary Tube with One-Way Valve (Ref . 281142) or Outflow Tubing
with One-Way Valve (Ref . 284649) .
11 . Insert the waste end of the outflow tubing into waste collection .
to end operation:
1 . Close the clamp under the pressure chamber . When another
surgery is scheduled for the same operating day, leave the
Inflow Tubing .
29