When there is no patient in the lift, wrap the leg
band around the leg support, close and shorten it
to tighten.

5.8 Emergency functions

5.8.1 Performing an emergency stop

1. Press the red emergency button A on the control unit
to stop the boom and patient from raising or lowering.
2. To reset, rotate the emergency button clockwise.
5.8.2 Activating an emergency lowering (CBJ
Home control unit)
If the hand control fails, the boom can be lowered by using
the circular switch for emergency release.
1. Lower the boom by pressing and holding button A at
the front of the control unit.
2. Stop lowering the boom by releasing the button.
5.8.3 Activating an emergency lowering (CBJ
Care, CBJ1, CBJ2 control unit)
If the hand control fails, the boom can be lowered by using
the circular switch for emergency release. Use a sharp
object, such as a pencil to press the button.
1. Lower the boom by pressing and holding button A at
the front of the control unit.
2. Stop lowering the boom by releasing the button.
1654371-A
5.8.4 Activating an emergency lifting (CBJ Care,
CBJ1 control unit)
If the hand control fails, the boom can be raised by using
the circular switch for emergency lifting. Use a sharp object,
such as a pencil to press the button.
1. Raise the boom by pressing and holding button B at the
front of the control unit.
2. Stop raising the boom by releasing the button.
5.8.5 Activating a mechanical emergency
lowering
If the emergency lowering on the control unit is not working,
a mechanical emergency lowering is available as a back-up.
This may happen in case of partial or total power failure, or
if the battery runs down during use.
WARNING!
Risk of Injury or Damage
Repeated mechanical emergency lowering leads
to a high increase of the lowering speed.
– Only use the mechanical emergency lowering
as a back-up when the emergency lowering on
the control unit is not working.
– Do not use the mechanical emergency lowering
repeatedly in close succession.
– Always lower the patient onto a supporting
surface like a bed or a chair.
– After a mechanical emergency lowering was
required, the lift must be checked to resolve
the malfunction of the control unit.
1. Locate the red
emergency grip A
at the bottom of the
actuator piston.
2. Slowly pull up the red
emergency grip A and
hold it in position when
a safe lowering speed
is reached.
3. If no lowering occurs
with a fully pulled
emergency grip A,
push down on the
boom at the same
time.
The effective lowering speed is depending on the patient
weight. If the speed is too low or too high, it can be adapted
to the patient weight.
Adjusting the lowering speed:
1. Locate the screw in the bottom hole of the red
emergency grip A.
2. Turn the screw counter-clockwise to increase the speed.
3. Turn the screw clockwise to decrease the speed.