6.2
For new units with Argovent 2
▷ Pull mains plug.
E N
▷ Unscrew the six screws on the rear of the fur-
nace and remove the back panel.
Fig. 12
▷ Mount rear insulating stones with the studs
from the delivery scope of the furnace.
▷ Insert (8 mm) floor plate from Argovent 2 deliv-
ery scope.
Fig. 13
28
▷ Thread the hose piece of the Argovent sintering
base through the hole of the rear wall and slide
the sintering base to the stop into the furnace;
if required, apply a light jiggling motion.
The hose piece must completely project out of
the rear side of the furnace.
Fig. 14
▷ Insert the free end of the supplied gas hose to
the internal gas connection.
Fig. 15