Filter and collection bottle
The hydrophobic filter retains water and aqueous solutions also in the form of aerosols. When using solvents, or due to
NOTICE
evaporation of water, formation of condensate at the outlet of the Aspirator is possible.
WARNING
Filter
Change filter if necessary in case of contamination, weak aspiration or no aspiration.
Note: After disinfection with alcohol, allow the filter to dry completely because the filter may
become clogged by alcohol.
Emptying the collection bottle of the Aspirator
Check liquid level in the collection bottle regularly.
Maximum admissible liquid level in collection bottle: approx. 80 %, depending on the application. (In case of low
boiling liquids or in case of suction of liquids which tend to foam, the admissible liquid level might be reduced.)
Switch off the Aspirator.
Vent the collection bottle.
Remove the connecting tube at the pump inlet, isolate connections in case.
Remove screw cap from the collection bottle after venting.
Remove bottle from the support.
Sterilize and dispose of collected liquid according to appropriate regulations.
14 · Aspirator · Manual
Important: Comply with applicable regulations when disposing of solvents.
Attention: Do not remove connections in case of liquid in the tube. Do not touch contaminated parts when removing
the screw cap. Risk of infection! Wear appropriate protective clothing, e.g., gloves.
U se only the original spare part (see section "Accessories - Spare parts") and sterilize and dispose of the filter
according to appropriate regulations.