3

Use and operation

3.1

First steps: Installation

3.2

Operating the Aspirator

The unit has a touch panel so the keys need only be touched gently. Do not press the keys.
The keys "+" and "-" have to be touched > 0.25 seconds to be actuated.
A successful action is confi rmed by a blip and the flashing of LEDs.
1
Use the keys to set suction power of the system.
The suction power can be set linearly in a range from 150 mbar (1 LED flashes) to 850 mbar (8 LEDs flash) underpressure (relative to
atmospheric pressure).
Depending on ambient atmospheric pressure (depending on altitude or weather conditions) and on the vapor pressure
NOTICE
of the media inside the collection bottle, the setting of a high underpressure (e.g., 850 mbar, 8 LEDs flash) may lead to
continuous operation of the pump. In this case, a reduction of the suction power is recommended.
12 · Aspirator · Manual
Unpack the equipment.
Read and comply with the document "Safety information for vacuum equipment"!
Assemble connection tubing with fi lter.
Setting up the Aspirator. Connect to power supply.
CAUTION
Check line voltage and current prior to switching on!
Attach aspiration hose of the hand controller to the hose nozzle of the bottle head.
Position
1
2
2
3
3
A flashing LED indicates that the actual available suction power differs from the preset suction
power.
Attention: If the collection bottle is under vacuum and the vacuum demand is reduced, the existing
underpressure inside the bottle remains until the vacuum is reduced (pressure rises) by further
aspiration with the hand controller.