Aim TTi TGR2050 Manuel D'instructions page 23

Masquer les pouces Voir aussi pour TGR2050:
Table des Matières

Publicité

Les langues disponibles
  • FR

Les langues disponibles

REF_OUT
REF_IN
REF_DIS
BUZZON
BUZZOFF
Editing and Cursor Movement Commands
FSTEP <nrf>
DBSTEP <nrf>
MVSTEP <nrf>
UVSTEP <nrf>
STEP_UP
STEP_DOWN
FIELD_UP
FIELD_DOWN
FREQ_PTR
LEV_PTR
MOD_TYPE_PTR
MOD_VAL_PTR
PKDEV_PTR
UTILS_PTR
STEP_PTR
System Commands
∗RST
∗CL <nrf>
RPP_RST
∗SAV <nrf>
22
Set the reference socket to output
Set the reference socket to input
Disable the reference socket
Enable the internal buzzer
Disable the internal buzzer
Set the frequency step size to <nrf> kHz
Set the dB step size to <nrf> dB
Set the linear step size to <nrf> mV
Set the linear step size to <nrf> uV
Performs the same function as pressing the ▲ key
Performs the same function as pressing the ▼ key
Performs the same function as pressing the FIELD ▲ key
Performs the same function as pressing the FIELD ▼ key
Moves the edit cursor to FREQUENCY and displays the appropriate menu to
make FREQUENCY viewable.
Moves the edit cursor to output LEVEL and displays the appropriate menu to
make output LEVEL viewable.
Moves the edit cursor to the MODULATION TYPE and displays the appropriate
menu to make MODULATION TYPE viewable.
Moves the edit cursor to the MODULATION VALUE and displays the
appropriate menu to make the MODULATION VALUE viewable.
Moves the edit cursor to PK DEVIATION and displays the appropriate menu to
make PK DEVIATION viewable.
Moves the edit cursor to the last selected parameter on the Utilities menu and
displays the Utilities menu.
Moves the edit cursor to the last selected parameter on the Step Size menu
and displays the Step Size menu.
Resets the instrument to default settings with the exception of all remote
interface settings.
Recalls the instrument set–up contained in store number <nrf>. Valid store
numbers are 1–10. Recalling store 10 sets all parameters to default settings
with the exception of remote interface settings. An attempt to recall from a
store which has not been previously loaded with a set–up will create an
execution error.
Reset the reverse power protection trip latch. If the reverse power is still
present the trip condition will be re-asserted immediately.
Saves the complete instrument set–up in store number <nrf>. Valid store
numbers are 1 – 9.