PROGRAMMING THE AUTOCLAVE

READ THIS PARAGRAPH CAREFULLY
BEFORE PROGRAMMING. ALWAYS CARRY OUT
PROGRAMMING WITH THE AUTOCLAVE DOOR OPEN.
To set the parameters shown on the display, proceed as follows:
Turn on the autoclave with the ON/OFF switch on the right side of the control panel, holding down the
START/STOP button; after a few seconds the Display shows the cycle counter; release the START/STOP
button. Press the START/STOP button to enter in the language setting menu.
With the TECHNICAL and SELECT buttons you can change the pre-selected language. The TECHNICAL
button increases values, the SELECT button decreases them.
Press the START/STOP button to go to the HOUR selection. The flashing number indicates the data you can
change. Press TECHNICAL and SELECT to change the current hour settings, bearing in mind TECHNICAL
will increase the value while SELECT will reduce it.
Press the DOOR button to pass to the MINUTE selection.
The flashing numbers are those you can change.
Press TECHNICAL and SELECT to change the minutes, bearing in mind that TECHNICAL increases the
number and SELECT reduces it.
Press the START/STOP button to pass to the DATE selection.
The flashing numbers show the value to be changed.
Press TECHNICAL and SELECT to change date settings, where TECHNICAL increases and SELECT
decreases the numbers.
Press the "DOOR" button to pass to the MONTH selection.
The flashing numbers show the value to be changed.
Press TECHNICAL and SELECT to change the month where TECHNICAL increases and SELECT
decreases the number.
Press the DOOR button to pass to the YEAR selection.
The flashing numbers show the values to be changed.
Press TECHNICAL and SELECT to change the year where TECHNICAL increases and SELECT decreases
the number.
Press the START/STOP button to match the atmospheric pressure. The display will show 0.00; keep the
door open until the pressure in the autoclave adjusts to the atmospheric pressure, which varies according to
the altitude. CARRY OUT THE ATMOSPHERIC PRESSURE SETTING ON FIRST INSTALLATION OF
THE AUTOCLAVE.
Press the DOOR button to confirm the input data.
Press the START/STOP button to save the settings. Keep pressing this button until you hear an acoustic
signal confirming all data are successfully stored.
The Display will show the selected sterilisation cycle.