PAR. 15

MAINTENANCE

CAUTION: ALWAYS DISCONNECT MAINS ELECTRICITY SUPPLY BEFORE CARRYING OUT
MAINTENANCE WORK
PAR. 15.1 - DAILY ROUTINE MAINTENANCE
CLEANING THE DOOR SEAL
- Clean the seal on the door and the outer edge of the sterilisation chamber with a wet cloth or the soft part
of the supplied sponge.
This cleaning operation must be carried out to remove any possible dirt which can cause a pressure loss or a
failure to obtain a vacuum in the sterilisation chamber, as well as damaging the seal itself.
PAR. 15.2 - WEEKLY ROUTINE MAINTENANCE
CLEANING THE STERILISATION CHAMBER
- To remove the deposits on the chamber floor, clean with the abrasive part of the supplied sponge after
removing the tray-holder.
- To moisten the sponge, only use distilled or desalted water.
PAR. 15.3 - QUARTERLY MAINTENANCE
- Lubricate the hinges and the closing pin with silicon oil.
- Replace bacteriological filter.
- Adjust door:
If the autoclave handle is too loose on closing, adjust as shown in the following pictures. Rotate anti-
clockwise (fig. 11) to increase the seal pressure on the chamber. If it has loosened too much and you are
unable to close the handle, turn it in a clockwise direction.
FIG. 11
FIG. 12
45